Table of modes of mechanical ventilation
In medicine, mechanical ventilation is a method to mechanically assist or replace spontaneous breathing. For this purpose, medical devices called Ventilators are used. Modern Ventilators offer a number of methods to deliver the breaths to the patient. These methods are called Modes of mechanical ventilation[1] (Mode) and are selected by the clinician. Common to all modes is that they allow the clinician to control
However, they differ in the way to deliver breaths[2]. Breath delivery follows the phases of breathing,[3] i.e., inhalation and exhalation:
CAVEAT: Although manufacturers may offer identical breath delivery methods, the names of the Modes may be different. Trigger: The ventilator needs to know when to start delivering gas to the patient. If the patient does not breathe at all, a timer starts inhalation. If the patient has some breathing activity, the ventilator can sense this effort by measuring pressure or flow and start inhalation if pressure or flow drop below a certain threshold. That threshold is called Trigger Sensitivity. Inhalation Mechanism: Technically, two methods to deliver the gas mixture can be employed, flow controlled or pressure controlled. Flow control means that the ventilator outputs a pre-set flow and maintains that flow until the end of inhalation. Pressure control means that the ventilator outputs a pre-set pressure and maintains that pressure until the end of inhalation. Both methods have their advantages and disadvantages. Flow control will deliver the gas mixture independent of resistance to flow and guarantee a set delivery of gas. In the process, pressure might become very high and potentially dangerous to the patient. Pressure control will deliver the gas mixture at a pre-set level and never exceed that pressure. However, it may not succeed to deliver a set volume of gas mixture. Cycling mechanism: Inhalation must eventually stop and enable to lungs to exhale. If the patient does not breathe, the ventilator must switch to exhalation after a pre-set time or after a pre-set volume has been delivered. If the patient has some breathing activity left, the ventilator can sense this by measuring flow and start exhalation, for example if flow drops below a certain threshold. That threshold may be termed "Expiratory Trigger Sensitivity". Exhalation Mechanism: Exhalation requires time for the lungs to empty. This time starts with the onset of exhalation and ends with the start of the subsequent inhalation. If the patient is passive, the exhalation is terminated by a timer. If the patient has some breathing activity, exhalation is terminated by the subsequent inhalation effort of the patient. The table below lists the working principles of some of the common modes of ventilation. (Vent = controlled by ventilator; Pat = controlled by patient, based on flow or pressure measurent)[4]. Some modes offer a special convenience feature, called Servo. For example: When Pressure Vontrolled Ventilation (PCV) is used, the volume delivered by the Ventilator may vary depending on the patient's lungs and the patient's efforts. Hypothetically, a clinician may sit next to the Ventilator and adjust the pressure control knob breath-by-breath to maintain a certain target volume. In Servo mode, the clinician inputs the target volume and the Ventilator adjusts the pressure control knob breath-by-breath.
See alsoReferences
|
Portal di Ensiklopedia Dunia